Brand Pipette Leak Testing UnitBrand Pipette Leak Testing Unit

Test which of your pipettes need servicing using the Brand Pipette Leak Testing Unit.

While the PLT (Pipette Leak Tester) unit cannot replace regular gravimetric testing, daily pipette checks can provide a safeguard during the periods between calibrations. Even the smallest leaks are detected! Process reliability for the pipettes is thus significantly improved.

The most frequent cause of inaccuracy in piston-operated pipettes is leakage. This arises from damage either to the seals, pistons, or tip cones. Often not detectable by the naked eye, leaks lead to significant volume errors.

The BRAND pipette leak tester (PLT unit) for air displacement pipettes finds even the smallest leaks within seconds.

  • Test limit values for all commercially available single and multichannel pipettes in the volume range 1l to 10ml are pre-programmed.
  • Test with and without tips to assist in leak source diagnosis .
  • Test results in seconds.


FURTHER DETAILS

Pipette Leak Testing Unit

Ref: BRA-703970
Quantity:



Price:1,040.00
(Excluding VAT at 20%)



Pipette Servicing
We offer a complete pipette maintenance and calibration service, which is available for most popular makes of pipettes.

Three service levels are offered to suit your needs - Bronze, Silver, and Gold.

In all the service options the pipette liquid end is disassembled, cleaned and visually inspected. The pipette is then reassembled, the accuracy and linearity are checked and, if required, adjusted to the manufacturer's specification.

New seals and ‘O’ rings will be fitted as standard. If a new seal and 'O' ring is not required, please specify at time of ordering. The price of parts is not included in the service price. Customers will be advised prior to replacement of any additional parts, other than seals and 'O' rings, if required.

The pipette service workshop and engineer both work to ISO 9002 standard and in accordance with DIN 12650. The calibration weights are traceable to UKAS accredited certification standard, and the software meets ISO 8655, DIN 12650, ISO 9000, NCCLS and GLP requirements. All service levels include a certificate of conformity

Most popular makes/models of pipette can be serviced. Turnaround time is generally within 3 days of receipt of the pipette (spare parts allowing)

SERVICE OPTION SINGLE CHANNEL CHARGE* MULTI CHANNEL CHARGE*
Bronze
(clean up plus 2:2 point certified calibration)
9.95/pipette33.00/pipette
Silver
(as bronze plus 2:4 point certified calibration)
15.95/pipette50.00/pipette
Gold
(as bronze plus 3:10 point certified calibration)
37.50/pipettePOA


All the above options are exclusive of any parts required, VAT and carriage.

A completed Decontamination Certificate is required for every pipette returned for servicing. Please click here to display one for printing.

Please send your pipettes direct to us at the following address - Thistle Scientific Ltd, DFDS House, Goldie Road, Uddingston, Glasgow, G71 6NZ along with an accompanying decontamination certificate, valid purchase order and specify the service level you require.